Slide Stainer Series

HS-2500 Fully Automated Immunohistochemical Stainer

1. Sample position: Equipped with 4 sets of dyeing units with a total of 48 dyeing positions,all of which can be started and stopped with one button.2. Independent temperature control: Each dyeing sta

1. Sample position: Equipped with 4 sets of dyeing units with a total of 48 dyeing positions,all of which can be started and stopped with one button.

2. Independent temperature control: Each dyeing station has independent temperature control, with an adjustment range of 35~105°C.

3. Reagent position: Equipped with 44 open reagent positions with refrigeration function, which can achieve 8~15°C temperature control adjustment.

4. Reagent Management: Reagent usage is visualized, and strict reagent management is achieved through barcode recording and liquid level detection.

5. Reagent information entry:1D code, 2D code.

6. Weighing function:Real-time monitoring of large-volume reagent volume through weighing function and free setting of alarm volume. 

7. Liquid adding method: The reagent needle can drip at multiple locations on the slide, and the volume of liquid added can be adjusted from 50-200ul.

8. Waste liquid collection: Toxic and hazardous waste liquids can be collected separately.

9. Software operation: The software can monitor the running status of each staining position in real time, and can customize the software settings and modify the experimental steps freely.

10. Machine processing speed: Fully loaded with 48 slides, 3 to 4 hours, complete the whole process of dewaxing, repairing and staining of 48 tissue slides.

11. Reagent compatibility:Compatible with mainstream reagents in the market, reagents can be managed in a closed manner.

12. Software Environment: Software installed on Windows.

13. Scope of application:The operating system supports multiple staining methods such as IHC, ISH, double staining, etc.

14. Sample Type:Applicable to paraffin sections, frozen sections, puncture specimens, cytology samples.

15. Information interface:RS 232 to USB interface.

16. Operating Environment:Temperature: 5-35"C, altitude 0-1600m, humidity (non-condensing) 30-80%RH.

17. Working power supply: Voltage: 220V±10%, Frequency: 50Hz

18. Instrument weight:About 135kg

19. Instrument size (length,width and height):1010mmx760mmx810mm